Abx Pentra 60 C+ User Manual

04 calibrator and control targets 01 abx pentra 400 calibrator target abx pentra hdl cal 13 Nov 2019 04 calibrator and control targets 01 abx pentra 400 calibrator target abx pentra apo cal 13 Nov 2019 04 calibrator and control targets 03 pentra c400 control target abx pentra. Horiba Medical - ABX Pentra 60Small, Convenient, Bench-Top Analyser with Excellent Analytical Performance for Efficient Patient Care, Highly. More ABX Pentra 60 C+ A Dedicated Workstation for a Powerful Management of your Workflow. More ABX Pentra DF 120 Productivity and reliability. More ABX Pentra DF 120 SPS Evolution Productivity and reliability. More ABX Pentra DX 120 In the heart of hematopoiesis.

  • Hematology
  • Data management on stand-alone PC.
  • Micro-sampling from whole blood (CBC: 30 µL - DIFF: 53 µL).
  • Cytochemistry, Impedance (real cell volume measurement) & Optical (analysis of the internal cellular structure by measuring light absorbance) - DHSS* Technology.
  • Perfect homogenization of blood samples with reagents - MDSS** Technology.

ABX Pentra 60 C+

  • Throughput: Up to 60 samples/hour.
  • Reagents: Only 4 onboard reagents and 1 diluent.
  • Perfect differentiation of the 5 WBC sub-populations with DHSS* Technology.
  • 3 histograms for RBC, BAS/WBC and PLT together with the 5 DIFF Matrix.
  • Basophils counted through specific channel.
  • High resolution matrix.

* DHSS: Double Hydrodynamic Sequential System

** MDSS: Multi-Distribution Sampling System

Manufactured by HORIBA ABX SAS

Technical

Weight

77 lbs / 35 kg

Power supply

100 to 240 V (± 10%) - 50 to 60 Hz

Dimensions

20.3 x 17.5 x 19 in / 51,5 cm x 44,5 cm x 48 cm

Printer

Laser printer

Throughput

60 tests/hour

Data management

Color screen: 15 in. monitor (800 x 600px)

Windows XP Prof.TM

User-defined flagging limits

Transmit patient & QC to LIS

Uni-directional or bi-directional connection

Communication Protocols: ABX / ASTM

CD-ROM drive / Floppy disk

2 serial ports minimum

USB Keyboard / USB Mouse

Memory

10,000 results + Graphics

Quality Control Management

12 selectable QC files

XB: 60 Operator selectable batches with statistics (20 samples per batch)

Within run

Levey-Jennings Graphs

Logs

Reagents, calibration, maintenance, errors, blank cycle


Samples

60 samples/hour (closed tube)

Parameters (26)

WBC, NEU# & NEU%, LYM# & LYM%, MON# & MON%, EOS# & EOS%, BAS# & BAS%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV

Graphs

High-definition LMNE matrix
PLT, RBC, WBC/BAS curves

Blood Volume

CBC mode: 30 µL
CBC + DIFF mode: 53 µL

Stability

48 hours post-draw stability

Operating sample mode

Closed tube

Identification

Alphanumeric. Bar-code reader (optional)

Results

Pentra

Parameters

Precision (%CV)

Linearity

WBC

< 2.0

0-120 x 10³/µL

RBC

< 2.0

0-8 x 106/µL

HGB

< 1.0

0-24 g/dl

HCT

< 2.0

0-67%

PLT (whole blood)

< 5.0

0-1900 x 10³/µL

PLT (PLT concentrate)

< 5.0

0-2800 x 10³/µL

One moment please...
  • Documents
  • Parts
  • Videos
  • News
DESCRIPTION
Small, Convenient, Bench-Top Analyser with Excellent Analytical Performance for Efficient Patient Care

  1. Highly compact and cost-effective.
  2. Micro-sampling from whole blood (CBC : 30 µL - DIFF: 53 µL)
  3. Cytochemistry, Impedance (real cell volume measurement) & Optical (analysis of the internal cellular structure by measuring light absorbance) - DHSS* Technology.
  4. Perfect homogenization of blood samples with reagents - MDSS** Technology.
FORUMSView All (17)
Ask a New Question

5

Replies
-TulioHPS
3 years ago
3 years agoPassword

Hello.

I need of the technician password to access the technician mode.

I will be grateful if someone send me.

Thanks.

Reply
-duyenhai
3 years ago
3 years agoHGB no test

My Pentra ES60 does not measure HGB (RBC, WBC, PLT be normal)?

Duyen Hai

Reply

0

Replies
-eliehraoui
4 years ago
4 years agomd

how to communicate with rs 232 of pentra 60

Reply
Pentra

FEATURES

  • Throughput: Up to 60 samples/hour

  • Reagents: Only 4 onboard reagents and 1 diluent

  • Perfect differentiation of the 5 WBC sub-populations with DHSS* Technology

  • 3 histograms for RBC, BAS/WBC and PLT together with the 5 DIFF Matrix.

  • Basophils counted through specific channel

  • High resolution matrix includes the determination of 2 additional subpopulations (% and #): Atypical Lymphocytes (ALY) and Large Immature Cells (LIC)

  • DHSS : Double Hydrodynamic Sequential System (HORIBA Medical Patent) ** MDSS : Multi-Distribution Sampling System (HORIBA Medical Patent) *** RUO parameters (Research Use Only)

SPECIFICATIONS

Weight77 lbs

Additional Specifications

Abx Pentra

Technical
Weight
77 lbs
Power supply
100/240 V - 50/60 Hz - 200 VA.
Dimensions
20 in x 18 in x 19 in.
Printer
dot matrix.
Throughout
60 samples/hour.
Data management
68 331 microprocessor.
4 counting channels (68HC11).
Mono-directional connection.
Memory
Last result.
Screen
16 lines LCD screen.
Samples
60 samples/hour (open tube)
Parameters (24+2)
WBC, RBC, PLT, HB, HCT, MCV, MCHC, MCH, PCT, MPV, PDW, RDW, % and # of LYM, MON, NEU, EOS, BAS, *ALY (# and %), *LIC (# and %).
Graphs
High-definition LMNE matrix
PLT, RBC, WBC/BAS
Volume
CBC mode : 30 µL
DIFF mode : 53 µL
Stability
48 hours
Operating sample mode
Open tube
Identification
Alphanumeric. Bar-code reader (optional)
Flags
46 possible pathologic flags.
* RUO (Research Use Only)